Please use this identifier to cite or link to this item: http://hdl.handle.net/123456789/2746
Title: All-reflective UV-Vis-NIR Absorption Spectroscopy setup for micro-sized samples
Authors: Patil, Parimal Pitambar
Keywords: Absorption spectroscopy
Spectrophotometry
Absorption Spectroscopyof Red Acrylic
Tungsten-Halogenlamp and Deuteriumlamp
Issue Date: 1-May-2024
Publisher: IISER- Mohali
Abstract: Absorption spectroscopy is the technique that measures the absorption of light incident on a material as a function of wavelength due to light-matter interactions. During the MS thesis, we set out to investigate whether a UV-Vis-NIR absorption spectroscopy setup can be built to analyse the samples in the range of a few micrometres. The setup was mesured for beam spot measurement for sample size to be less 20 micrometres. The setup uses all-reflective components in its design to minimise spectral losses. We demonstrate the capability of the setup using common spectrally active materials like red-dyed acrylic, gel ink and yellow volatile corrosion inhibitor (VCI) film. The performance is validated using a commercial UV-Vis-NIR absorption spectroscope.
URI: http://hdl.handle.net/123456789/2746
Appears in Collections:MS-19

Files in This Item:
File Description SizeFormat 
Need To Add…Full Text_PDF (1) (3).pdf19.04 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.